What is a recommended method for using the Sep-Pak Accell Plus QMA to retain and elute an acidic compound? - WKB124677
Article number: 124677
ENVIRONMENT
- Sep-Pak Accell Plus QMA
- Method development
- Acid retention
ANSWER
- Condition your SPE Classic cartridge with 6 -10 hold-up volumes of 100% water.
- Equilibrate with 6-10 hold-up volumes of your loading conditions (experiment with pHs 5-8 for best ion exchange of target acid).
- Load sample (1-5 mL volume is typical but can be higher if sorbent is not overloaded).
The flow should look like droplets (not too fast; maximize contact time with the sorbent); 1 drop per second flow rate.
- Wash with 3-6 hold-up volumes of buffer pH 5-8 to reinforce the ion exchange of target acid to sorbent.
- Elute with acidic water (perhaps 5% acetic acid or 1% formic acid) or some combination of acidic water and 50% methanol or acetonitrile.
The recommended elution step volume is 3 cartridge hold-up volumes (you can cut this in half if a reasonable recovery is maximized by optimized elution conditions).
The elution flow should look like droplets (not too fast; maximize contact time with the sorbent); 1 drop per second flow rate.
ADDITIONAL INFORMATION
For additional information, see the Waters AccellPlus QMA and CM Bulk Media Care and Use Manual.