Can I use the 96-Flangeless SPE Cartridge Holder (186005523) on a 96-well plate vacuum manifold? - WKB122720

Article number: 122720

ENVIRONMENT

  • Solid-phase extraction (SPE)
  • 96-well plates
  • Flangeless SPE Cartridge Holder
  • Vacuum manifold

ANSWER

No

ADDITIONAL INFORMATION

There is no seal between the bottom of the flangeless cartridges and the 96-Flangeless SPE Cartridge Holder. A proper vacuum will not form to pull liquid through the cartridges.