How do I use MS cleaning solution? - WKB5736
Article number: 5736
OBJECTIVE or GOAL
Cleaning with MS cleaning solution
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-MS
- Xevo TQ-D
- SQD2
- TQD
- SQD
- Quattro Premier
- Quattro Micro
- SYNAPT
- Xevo G2-S
- Xevo G2-XS
- Xevo TQ-XS
- Vion IMS QTof
- Xevo G3 QTof
PROCEDURE
- Place the part you want to clean in a appropriately sized beaker.
- Pour MS cleaning solution (186006846) into the beaker until the item is covered.
- Sonicate* for 20-30 minutes.
- Dump out the MS cleaning solution into an appropriate waste container.
- Rinse the item with water until all suds are gone.
- Sonicate in high-quality water for 10 minutes.
- Dump out the water and replace it with high-quality isopropanol or methanol.
- Sonicate for 20-30 minutes.
- Dry the item with an oil-free nitrogen supply.
ADDITIONAL INFORMATION
*If the ultrasonic bath has a heating function it is recommended to heat the cleaning solution while sonicating. Heating 60 to 70°C improves the efficiency of cleaning. Do not use the ultrasonic bath heating function when using isopropanol or methanol (Fire hazard)!
Which chemicals are in MS Cleaning Solution (186006846) - WKB107252