How do I prepare mobile phases for HILIC applications? - WKB31055

Article number: 31055

OBJECTIVE or GOAL

Prepare mobile phase for HILIC applications.

ENVIRONMENT

  • Hydrophilic Interaction Chromatography (HILIC)
  • Mobile phase preparation

PROCEDURE

Low pH

Prepare a 200 mM stock buffer of ammonium formate by dissolving 12.6 g of formic acid ammonium salt and adding it to 900 mL of water in a 1 L volumetric flask. Twenty five milliliters of formic acid was added. Water was then added to the 1 L mark. Final solution pH was 3.0.

If using a Binary Solvent Manager system, prepare the low pH A and B mobile phases as follows:

  • Mobile phase A: 90/10 ACN/water with 10 mM ammonium formate and 0.125% formic acid: Add 50 mL of 200 mM stock buffer and 50 mL of MilliQ water to 900 mL of ACN.  Sonicate in warm water bath to eliminate cloudiness.
  • Mobile phase B: 50/50 ACN/water with 10 mM ammonium formate and 0.125% formic acid: Add 50 mL of 200 mM stock buffer and 450 mL of water to 500 mL of ACN. 
High pH

Prepare a 200 mM stock buffer of ammonium acetate by dissolving 15.4 g of ammonium acetate and adding it to 900 mL of water in a 1 L volumetric flask. Eight milliliters of ammonium hydroxide was added. Water was then added to the 1 L mark. The final solution pH was 9.0.

 If using a Binary Solvent Manager system, prepare the high pH A and B mobile phases as follows:

  • Mobile phase A: 90/10 ACN/water with 10 mM ammonium acetate and 0.04% ammonium hydroxide: Add 50 mL of 200 mM stock buffer and 50 mL of MilliQ water to 900 mL of ACN. 
  • Mobile phase B: 50/50 ACN/water with 10 mM ammonium acetate and 0.04% ammonium hydroxide: Add 50 mL of 200 mM stock buffer and 450 mL of water to 500 mL of ACN.