Do I need both the UV and the fluorescence detector hooked up to an ACQUITY system for AccQ•Tag Ultra derivatized analysis of tryptophan? - WKB125064

Article number: 125064

ENVIRONMENT

  • AccQ•Tag Ultra
  • ACQUITY H-Class
  • UPLC Classic
  • Tryptophan
  • Amino acid analysis (AAA)
  • UV detection
  • Fluorescence detection

ANSWER

No. When using ACQUITY UPLC or H-Class systems, only a UV detector is needed for analysis of AccQ•Tag-derivatized tryptophan.

ADDITIONAL INFORMATION

For additional information, see Enhancement of the UPLC Amino Acid Analysis Solution with Flexible Detector Options.