Skip to main content
Waters

Error: Database failure opening chromatogram for write - WKB122253

Article number: 122253

SYMPTOMS

  • Error message "Database failure opening chromatogram for write" appears when trying to start a sample set or a single injection

ENVIRONMENT

  • Empower 3

CAUSE

Project file is corrupted.

FIX or WORKAROUND

Create a new project file.

ADDITIONAL INFORMATION

Similar error information: WKB64508

Not able to find a solution? Click here to request help.