What is the shelf life or expiration point of the consumable chemistry components in the HPLC AccQ•Tag chemistry kit (WAT052875)? - WKB119392

Article number: 119392

ENVIRONMENT

  • AccQ•Tag for HPLC
  • Amino acid analysis (AAA)
  • AccQ•Fluor
  • Standards

ANSWER

WAT052880 - ACCQ-FLUOR REAGENT PKG TESTED :

  • 1 year expiration from the date of manufacture
  • Store at room temperature when unopened until expiration date on label.
  • Reconstituted AccQ•Fluor Reagent can be stored at room temperature in a desiccator with dry desiccant for up to one week.
  • Derivatized amino acid samples may be stored at room temperature for up to one week.

 

WAT052885 - ACCQ-TAG 3.9X150mm Column:

  • This column does not have an expiration date.
  • Waters recommends that each laboratory assign expiration dating based on performance of the column for the method in use.
  • Store at room temperature.

 

WAT052890 - ACCQ-TAG ELUENT A CONCENTRATE - 1 Liter Bottle:

  • Store at room temperature when unopened until expiration date on label.
  • After opening a container of AccQ•Tag Ultra eluent A concentrate, the eluent remains serviceable for 1 month at room temperature.
  • Stored at 4 °C, tightly capped in original container, it remains serviceable for one month.

 

WAT088122 – Amino Acid Standard:

  • 5 year expiration from the date of manufacture
  • Unopened ampoules or aliquots of this standard should be stored at -20 °C.  
  • Once removed from the freezer, allow to sit at room temperature until it is ready to use.
  • Do not thaw with a hot plate and do not thaw with heat from hands.
  • Once opened, use immediately. 