What might be the source of peak fronting that is visible on the Alliance 2695 but not on the Agilent 1100 or 1290? - WKB119278
Article number: 119278
ENVIRONMENT
- Alliance 2695
- Agilent 1100
- Agilent 1290
- Peak fronting
ANSWER
The way in which the injector needle is cleaned is different in the Alliance 2695 versus Agilent systems.
ADDITIONAL INFORMATION
The FTN of the Alliance 2695 has a needle purge feature and a separate wiper/frit wash. The needle purge wash is injected along with the sample. This should match the starting conditions of the mobile phase. If the mobile phase contains 30% organic but the needle purge wash is 50% organic, strong solvent effect would cause peak shape problems. The wiper/frit wash can be a stronger wash solution composition, because this only cleans the outside of the injector needle.