How do I reduce injector contamination when running the UPLC amino acid total solution? - WKB117451
Article number: 117451
OBJECTIVE or GOAL
Reduce or eliminate fouling of injection needle and injector loop.
ENVIRONMENT
- ACQUITY UPLC
- Amino acid analysis
- AccQ•Tag Ultra
- Injector loop
- Injection needle
PROCEDURE
- Each time you refresh the washes, you should replace all containers for the seal washes and strong and weak needle wash with new, clean containers and fresh solvent. Do not reuse the container or top off the liquid in the container.
- Weak needle wash should be remade in a clean container every five days.
- Seal wash and strong needle wash should be remade and placed in clean containers every two weeks.
ADDITIONAL INFORMATION
For additional information, see How to clean and flush the Alliance HPLC System and ACQUITY UPLC System to remove contamination and What should I do to prevent chromatographic contamination in the amino acid kits like Accq-Tag and Pico-Tag?
