Can I store a BEH SEC column in 80:20 Water:Methanol in a refrigerator at 2-8 °C? - WKB98518
Article number: 98518
ENVIRONMENT
- UltraPerformance Liquid Chromatography (UPLC)
- Ultra High Performance Liquid Chromatography (UHPLC)
- High Performance Liquid Chromatography (HPLC)
- Column
- ACQUITY
- XBridge
- Bridged Ethylene Hybrid (BEH)
- Size Exclusion Chromatography (SEC)
- Protein Separation
- Column Storage
ANSWER
Yes.
ADDITIONAL INFORMATION
You can store BEH SEC columns in a refrigerator (2-8 ºC) when they contain 80:20 Water:Methanol, although it is not required. The 20% methanol inhibits microbial growth.
ACQUITY UPLC BEH SEC Columns and Standards Care & Use Manual
XBridge Protein BEH SEC Columns & Standards Care & Use Manual
Guide to Size-Exclusion Chromatography (SEC) of mAb Aggregates, Monomers, and Fragments