Skip to main content
Waters

Can I store a BEH SEC column in 80:20 Water:Methanol in a refrigerator at 2-8 °C? - WKB98518

Article number: 98518

ENVIRONMENT

  • UltraPerformance Liquid Chromatography (UPLC)
  • Ultra High Performance Liquid Chromatography (UHPLC)
  • High Performance Liquid Chromatography (HPLC)
  • Column
  • ACQUITY
  • XBridge
  • Bridged Ethylene Hybrid (BEH)
  • Size Exclusion Chromatography (SEC)
  • Protein Separation
  • Column Storage

ANSWER

Yes.

ADDITIONAL INFORMATION

You can store BEH SEC columns in a refrigerator (2-8 ºC) when they contain 80:20 Water:Methanol, although it is not required. The 20% methanol inhibits microbial growth.

ACQUITY UPLC BEH SEC Columns  and Standards Care & Use Manual

XBridge Protein BEH SEC Columns & Standards Care & Use Manual

Guide to Size-Exclusion Chromatography (SEC) of mAb Aggregates, Monomers, and Fragments

id98518, monoclonal antibody, UPBINARY

Not able to find a solution? Click here to request help.