Is it OK to inject a sample that contains 2-molar ammonium sulfate onto 186004505 (XBridge Protein BEH C4 column, 300Å, 3.5 µm, 4.6 mm X 250 mm)? - WKB91618

Article number: 91618

ENVIRONMENT

  • XBridge Protein BEH C4 column

ANSWER

Yes, provided solubility of 2-molar ammonium sulfate in the initial mobile phase is considered.

ADDITIONAL INFORMATION

Before injecting a sample with high salt concentration, it is important to verify that the sample is completely soluble in your initial mobile phase, so that the high salt does not precipitate upon injection and cause column problems.