ACQUITY UPLC high delta pressure - WKB88253

Article number: 88253

SYMPTOMS

Delta pressure > 150 psi

ENVIRONMENT

  • H-Class
  • I-Class
  • ACQUITY UPLC
  • ACQUITY Arc

CAUSE

Clogged filter in the ACQUITY Column in-line filter (205000343).

FIX or WORKAROUND

Replace the filter (700003776).

ADDITIONAL INFORMATION

This can also occur when the user's column has degraded.