Is it OK to backflush an ACQUITY UPLC BEH Amide column? - WKB85794
Article number: 85794
ENVIRONMENT
- UPLC
- Backflush
- ACQUITY BEH Amide
ANSWER
Yes, at a slow flow rate
ADDITIONAL INFORMATION
When backflushing an ACQUITY UPLC BEH Amide column:
- Disconnect the guard column if you are using one.
- Disconnect the column and install it in the reverse direction.
- Do not connect the column to your detector.
- Flow to waste.
- Set the flow rate to 0.1 mL/min for 2.1-mm ID columns or 0.3 mL/min for 3.0-mm ID columns.
See also How do I clean an ACQUITY BEH Amide or XBridge Amide column?
id85794, UPBINARY