How do I clean an ACQUITY BEH Amide or XBridge Amide column? - WKB71990
ENVIRONMENT
- ACQUITY
- clean
- BEH
- XBridge
- Amide
- columns
- purge
- care
ANSWER
To clean polar compounds from the column, run a 10-minute gradient from 0% to 100% water.
Note that back pressure will increase as the water concentration increases.
Increasing temperature can help clean the column.
The BEH Amide's temperature limit is 90 °C, and the pH range is 2-11.
ADDITIONAL INFORMATION
ACQUITY and ACQUITY Premier Columns Care and Use Manual
XBridge and XBridge Premier Columns Care and Use Manual
NOTE: Carbohydrates, simple sugars, and polysaccharides are not soluble in >80% organic. When cleaning a column exposed to these types of samples or matrices with sugars and high salts, the cleaning medium should not exceed 80% organic—to avoid precipitation of these types of compounds. Tips for carbohydrates are included in the following:
Section VI d, "Tips for Separating Sugars/Saccharides/Carbohydrates", in the ACQUITY BEH columns care and use manual
Section VII f, "Tips for Separating Sugars/Saccharides/Carbohydrates", in the XBridge BEH columns care and use manual