What flow rate is safe to convert the IC-Pak Ion Exclusion column to 10% methanol for storage? - WKB75094

Article number: 75094

ENVIRONMENT

  • IC-Pak Ion Exclusion column
  • Storage solvent

ANSWER

The IC Pak Ion Exclusion columns should be flowed at 0.1–0.2 mL/min to minimize stress to the sorbent as the column moves from 100% aqueous to 10:90 methanol/water.

ADDITIONAL INFORMATION

 