What pore size packing should I use to protect my ACQUITY APC columns? - WKB72525
Article number: 72525
ENVIRONMENT
- Advanced Polymer Chromatography (APC)
- ACQUITY APC columns
- Guard columns
ANSWER
Waters recommends that a guard column contain the same particles that are packed in the first column of the column bank.
ADDITIONAL INFORMATION
For example, if the first ACQUITY APC column in a column bank has an APC XT 450, 2.5-µm packing, the guard should have the same APC XT 450, 2.5-µm packing.
ACQUITY APC Columns Care & Use Manual