Is the ACQUITY CSH C18 column compatible with 100% aqueous mobile phase? - WKB68339
Article number: 68339
ENVIRONMENT
- ACQUITY CSH C18 column
- Aqueous mobile phase
- water
- compatible
- care and use
ANSWER
Yes. However, the column should be exposed to at least 20% organic every day of use.
ADDITIONAL INFORMATION
For additional information, see the ACQUITY UPLC CSH Columns Care and Use Manual.