What is the recommended injection volume for an ACQUITY UPLC column? - WKB66208
Article number: 66208
ENVIRONMENT
- ACQUITY UPLC columns
ANSWER
The recommended injection volume for an ACQUITY UPLC column is 1-3% of the total empty column volume.
ADDITIONAL INFORMATION
Ultimately, the response of a particular sample in a particular detector may dictate the injection volume used to keep the response in the linear range.
For additional information, see How do I calculate the empty column volume of a column?