How do I determine the flow rate scaling from one preparative column ID to another? - WKB56490

Article number: 56490

ENVIRONMENT

  • Preparative column
  • Flow rate scaling

ANSWER

Use the Waters online preparative columns calculator.

Online Preparative Columns Calculator: http://www.waters.com/waters/promotionDetail.htm?id=134713448

ADDITIONAL INFORMATION

The following equation can be used to determine scaing from one column ID to another:

Flow rate(prep) = Flow rate(analytical) x [(diameter prep)2/(diameter analytical)2] x [particle size(analytical)/particle size(prep)]

Not able to find a solution? Click here to request help.