What is the long-term storage solvent for the Carbohydrate Analysis column (WAT084038)? - WKB226887

Article number: 226887

ENVIRONMENT

  • High Performance Liquid Chromatography (HPLC)
  • Hydrophilic interaction chromatography (HILIC)
  • Column care and use
  • Carbohydrate Analysis

ANSWER

100% acetonitrile

ADDITIONAL INFORMATION

For periods of three days or less, store this column in 80:20 water/acetonitrile (or similar HILIC mobile phase).

For periods of more than three days, store this column in 100% acetonitrile.

Note:

If this column is kept in the presence of water or methanol for long-term storage, the NH2 ligands can hydrolyze off of the particles, which can lead to selectivity changes.

Also see:

Retention Times for Sugar and Carbohydrates

≥3 µm Analytical HPLC Columns Chapter from current catalog