What LC fittings are recommended for the inlet and outlet of the HSS T3 column? - WKB204714
ENVIRONMENT
- Column fittings
- HPLC column hardware
- ACQUITY UPLC column hardware
- XP column hardware
ANSWER
If the HSS T3 is packed in ACQUITY UPLC hardware or is a 2.5-µm XP column, see What are the part numbers for the fittings that connect an ACQUITY UPLC column or a 2.5-µm eXtended Performance (XP) Column to a non-Waters UPLC system? and What are the part numbers for the fittings that connect an ACQUITY UPLC column or a 2.5-µm eXtended Performance (XP) column to an ACQUITY system that does not use an active column preheater?
If the HSS T3 is packed in HPLC hardware, use the following parts:
186008714 - PEEK Finger-tight Fitting, 10-32, 1 each, for inlet and outlet
or
WAT025604 - FITTING SS 5/PK (contains five stainless steel compression screws and five stainless steel ferrules) for inlet and outlet
ADDITIONAL INFORMATION
For additional information, see How do I properly connect the XSelect HSS HPLC columns? and How do I connect an ACQUITY column using the Finger-Tight Reusable Fitting Kit (700003139)? and How do I install a column onto the Active Preheater on an ACQUITY Arc System?
id204714, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, APH, UPBINARY