How do I clean my µBondapak Cyano (CN) column? - WKB123612
Article number: 123612
OBJECTIVE or GOAL
Clean a µBondapak Cyano column.
ENVIRONMENT
- µBondapak Cyano
- Column care and cleaning
PROCEDURE
- Remove the guard column.
- Prepare a solution of 50:50 dimethylsulfoxide (DMSO)/water.
- Disconnect the column outlet from the detector and catch flow in a beaker.
- Run 100% water as the mobile phase.
- Make 200-µL injections every 5 minutes until a total of 10 column volumes of water has passed through the column.
- After cleaning, connect the guard.
- Connect the column to the detector.
- Equilibrate the column with working mobile phase.
- Make an injection to check performance.
ADDITIONAL INFORMATION
If the cleaning above does not correct the issue, you can try a column regeneration procedure:
- Remove the guard column.
- Disconnect the column outlet from the detector and catch flow in a beaker.
- Run 10 column volumes of each of the following solvents through the column:
- Water
- Methanol
- Chloroform
- Methanol
- Water
Note: The formula to calculate your column volume is: pi * r2 * L (r = radius in cm, L = length in cm)