How do I reduce the back pressure in a method that is being developed at a high flow rate? - WKB118861
Article number: 118861
OBJECTIVE or GOAL
Reduce the back pressure of an LC method and stay within the column pressure limits.
ENVIRONMENT
- Method development
- Column selection
PROCEDURE
In order to run a column within its manufacturer's suggested pressure ranges, one or several of the following options would be implemented.
- Use a shorter column length.
- Use a wider column ID.
- Use a larger particle size.
- Raise the column temperature.
- Reduce the flow rate
- Consider using larger ID tubing post-column to reduce the friction produced by the method running at very high flow rates.