Skip to main content
Waters

In what order should I place the GPC columns and why? - WKB11879

Article number: 11879

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • GPC columns
  • Styragel

ANSWER

Place 100A or 50A columns (smaller pore size columns) at the end of the set, because the gel in these columns tends to be softer and less durable.

Not able to find a solution? Click here to request help.