How long will it take my AccQ•Tag Ultra column to equilibrate? - WKB117479
Article number: 117479
ENVIRONMENT
- AccQ•Tag Ultra
- Column equilibration
ANSWER
The AccQ•Tag Ultra column may take three to six standard injections to fully equilibrate out of storage or brand new.
ADDITIONAL INFORMATION
Use three derivatization blank injections followed by three standard injections to start equilibration. This has always been the case with reverse-phase amino acid methods.
id117479, 2695UP