Reduced digestion efficiency on new Enzymate BEH Pepsin column attached to HDX system - WKB117306

Article number: 117306

SYMPTOMS

  • New Enzymate BEH Pepsin column is not digesting Phos B standard efficiently and is giving very low recovery of expected peptides at startup

ENVIRONMENT

  • HDX system
  • Enzymate BEH Pepsin column
  • HDX Phosphorylase B standard

CAUSE

Quenching buffer reagents are going bad.

FIX or WORKAROUND

  1. Check Enzymate column efficiency with the Waters HDX Phosphorylase B standard {186006930) and compare peptide digestion recovery.  
  2. Quench labeling reaction with an alternative buffer.
  3. Remake method quenching buffer from a fresh lot of reagents.

ADDITIONAL INFORMATION

For additional information on using the HDX Phos B standard for troubleshooting, see the Enzymate BEH Pepsin Columns Care and Use Manual.