What are recommended cleaning procedures for Nova-Pak Phenyl columns? - WKB111949
Article number: 111949
ENVIRONMENT
- Nova-Pak Phenyl columns
- Column care and cleaning
ANSWER
If more polar samples were injected, the cleaning solvents would be water, methanol, THF, methanol, water, and mobile phase.
If more nonpolar samples were injected, the cleaning solvents would be IPA (or IPA/water mix), THF, DCM, hexane, IPA, and mobile phase.
ADDITIONAL INFORMATION
For additional information, see What are general tips for column cleaning? and Analytical Columns and Cartridges Care and Use Manual.