Can an ACQUITY UPLC BEH300 C4 1.7µm 2.1 x 50mm column (186004495) be run in an HPLC system that has maximum pressure of only 400 bar (5,801 psi)? - WKB111609

Article number: 111609

ENVIRONMENT

 

  • UltraPerformance Liquid Chromatography (UPLC)
  • Columns
  • ACQUITY
  • Bridged ethylene hybrid (BEH)

 

ANSWER

This depends on the method conditions.

ADDITIONAL INFORMATION

Enter the method conditions on our Columns Calculator Version 2.0 to determine what the maximum estimated back pressure would be on this column under the given method conditions. If the estimated maximum back pressure is less than 400 bar (5,801 psi), this column can be run on the system. Note that the maximum pressure value may not include additional system pressure.