How to perform a solvent changeover procedure on Waters Styragel columns to convert from DMF to NMP - WKB110513
Article number: 110513
OBJECTIVE or GOAL
Convert Styragel columns packed in dimethylformamide (DMF) to N-methyl-2-pyrrolidone (NMP)
ENVIRONMENT
- Gel permeation chromatography (GPC)
- Styragel
- Solvent changeover
- DMF
- NMP
PROCEDURE
- Set the column oven to 50–65 °C.
- For 7.8 x 300 mm columns, set the pump flow rate to 0.1 mL/min throughout this procedure.
- For 6.0 x 150 mm columns, set the pump flow rate to 0.06 mL/min throughout this procedure.
- For 4.6 x 300 mm columns, set the pump flow rate to 0.03 mL/min throughout this procedure.
- If you are running a gradient, program a 95:5 DMF/NMP to 100% NMP gradient over four hours.
- If you are running in Isocratic mode, make separate DMF:NMP solutions in 5% steps from 95:5 DMF/NMP to 100% NMP, and increase NMP strength over four hours.
- Flush with 100% NMP for six hours per column. If there are two columns in the column bank, flush for 12 hours. This equates to 10 column volumes per column.
-
This solvent switch should be performed only once in the column's lifetime.
ADDITIONAL INFORMATION
GPC Columns – Operating Procedures:
- Always ramp from 0.1 mL/min to approximately 1 mL/min over five minutes, for 7.8 x 300 mm columns, for example.
- Do likewise when ramping down flow rate.
- Typical operating flow rates when performing GPC separations on Styragel columns:
- 7.8 x 300 mm columns - 1.0 mL/min
- 6.0 x 150 mm columns - 0.6 mL/min
- 4.6 x 300 mm columns - 0.3 mL/min
- When the column is not in use but is kept on a system, keep the flow rate at 0.1 mL/min minimum at all times. Do not power-off the pump.