How to prepare the 200 pg/μl LeuEnk lockmass solutions for Xevo G2-S/XS or SYNAPT G2-S - WKB93406

Article number: 93406

OBJECTIVE or GOAL

Prepare the lockmass solutions for the Xevo QTof family and the SYNAPT family.

ENVIRONMENT

  • Xevo QTof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-S QTof
  • Xevo G2-S Tof
  • Xevo G2-XS
  • Xevo G2-XS QTof
  • Xevo G2-XS Tof
  • Xevo G3 QTof
  • SYNAPT G2
  • SYNAPT G2 MS/HDMS
  • SYNAPT G2-S MS/HDMS
  • SYNAPT G2-Si MS/HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si

PROCEDURE

The lockmass solution is leucine enkephalin (LeuEnk) prepared at a concentration of 200 pg/μl.

  1. Prepare a stock at 1 mg/mL (μg/μL) of LeuEnk solution by adding 3 ml of water to the Waters reference standard #186006013. Do not use organic solvent, because the leucine enkephalin will not dissolve.
  2. Add 100 μL of the LeuEnk stock solution to 500 mL of 50:50 water/acetonitrile + 0.1% formic acid.