How to make sodium formate from the sodium hydroxide solution provided in Tof G2-S Sample Kit-2 - WKB10562
Article number: 10562
OBJECTIVE or GOAL
Make sodium formate from the 0.5 mM sodium hydroxide provided in Tof G2-S Sample Kit -2 (700008892)
ENVIRONMENT
- Tof G2-S calibration solution
- Sodium formate calibration solution
PROCEDURE
- Using a 1000-μL pipette, transfer 900 μL of deionized or LCMS-grade water to the 5-mL vial.
- Using a 200-μL pipette, add 100 μL formic acid to the 5-mL vial and shake well.
- Using a 200-μL pipette, transfer 100 μL of the 10% formic/aqueous solution to the 30-mL nalgene bottle containing the 0.5 mM sodium hydroxide; sonicate for 5 minutes.
- Label the bottle “0.5 mM sodium formate”; store in a refrigerator for up to 3 months.