Mounting the µMALDI source
Article number: 224528
To mount the µMALDI source:
- Vent the instrument.
- Switch off the electronics.
WARNING: To avoid burn injury, wait at least 30 minutes for the ion block to cool before removing it. |
- Remove the ion block.
- Fit the µMALDI mounting flange to the instrument pumping block.
- Raise the µMALDI source and align it with the mounting flange.
- Push the µMALDI source onto the locating dowels.
- Clamp the µMALDI source into position.
- Connect all cables.
- Switch on the electronics.
- Pump down the instrument.