Connecting the nitrogen gas supply
Article number: 208165
Required materials
- Chemical-resistant, powder-free gloves
- 6-mm PTFE tubing (included in the SELECT SERIES MRT installation kit)
- Nitrogen regulator (not supplied)
To connect the nitrogen gas supply
- Connect one free end of the 6-mm PTFE tubing to the API gas port on the rear of the instrument.
Requirement: Do not cut the 6-mm (1/4-inch) PTFE tubing to size. Use the entire 5-m (16-ft) length supplied.
Figure 1: SELECT SERIES MRT gas and vacuum connectors
|
|
|
|
|
|
|
|
|
|
|
|
- Attach a nitrogen regulator to the nitrogen supply.
Requirement: The nitrogen must be dry and oil-free with a purity of at least 95%.
- Connect the free end of the 6-mm PTFE tubing to the nitrogen regulator.
- Set the nitrogen regulator to 6.5 to 7.0 bar (94 to 102 psi).
- Ensure that there are no gas leaks at any of the nitrogen gas supply fittings.