To select REM mode:

  1. Open MassLynx, and then select the Instrument tab.
  2. Click the MS Tune icon.

clipboard_e4a453fc704f405bd6a2ae69dd241d9ed.png

  1. From the Tune page window, click the REM button.

clipboard_e3523a69f953196a57495202f6d347db5.png

  1. Optimize the Transfer RF for the m/z of interest.

NOTE:    The following table lists the recommended settings for the corresponding m/z ranges (Transfer RF gain = 1)

m/z Range Transfer RF
50 - 150 75
100 - 300 150
200 - 600 400
300 - 1000 500
600 - 2000 800

 