Venting and powering-off the instrument for routine maintenance

Article number: 236170

IMPORTANT:

  • Venting and powering-off the instrument is a task best performed by advanced users or maintenance personnel. Attempt this operation only if you understand the consequences of errors committed while doing so.
  • If you intend to power-off the instrument, vent the instrument before you discontinue power.

To vent the instrument:

To vent the instrument, from the navigation menu, if the instrument is in Operate mode, click Standby.jpg (Source Standby), and then click Vent instrument. When a pump or vent command is issued, the status of the operation updates in real time.

To power-off the instrument once vented:

To power-off the mass spectrometer once vented, disconnect the power cable from the electrical outlet.

Not able to find a solution? Click here to request help.