Skip to main content
Waters

205000609 Waters ACQUITY Fluorescence Detector Large Volume Flow Cell Option Kit

Article number: 274126

ENVIRONMENT

  • ACQUITY UPLC FLR Detector

CONTENT

Waters ACQUITY Fluorescence Detector Large Volume Flow Cell Option Kit - Part Number 205000609

Part Number Quantity Item Image Description/Location

 

 

289004509

 

 

1

 

 

ASSY, FLOW CELL, LARGE VOLUME FLR

289004509.JPG

 

 

Flow Cell

 

289004745

 

1

 

REGULATOR, BACK PRESSURE, 100 PSI

 289004745.jpg

 

Back pressure regulator connects to the oultet of the flow cell

430001892 1 ASSY, TUBE INLET .0025 ID PEEK NUT FLR 430001892.JPG Inlet tubing to the flow cell

ADDITIONAL INFORMATION

Installation requirements:

When installing the large-volume fluorescence flow cell (part number 289004509) on the FLR detector, use the pressure regulator (part number 289004745) included in the Large Volume FLR Flow Cell Kit (part number 205000609).

CAUTION: Do not use the standard pressure regulator shipped with the FLR detector. That regulator’s 250-psi set point, which exceeds the large-volume flow cell’s 145-psi pressure limit, can damage the flow cell.

Refer to the ACQUITY UPLC System FLR Detector Getting Started Guide (part number 71500142403) for instructions on installing the flow cell and pressure regulator.

Not able to find a solution? Click here to request help.