How to enable FractionLynx to start the collection at a specific tube in the collection rack - WKB7453
Article number: 7453
OBJECTIVE or GOAL
Enable FractionLynx to start the collection at a specific tube in the collection rack.
ENVIRONMENT
- Waters 2767 Sample Manager
- Waters 2757 Fraction Collector
- Waters Fraction Collector 3
PROCEDURE
- Under the FractionLynx shortcut, open Collection Control.
- In Collection Control, open Collector Setup.
- In Collector Setup, click on System.
- Under System, click on Collection.
- In the Sequential section, enable Sequential Collection and select "Enable User Start Location".
- Save the settings and load them to the fraction collector.
- In the sample list, add the column Fraction Start.
- In the Fraction Start field, enter the tube location (using the appropriate syntax) where the collection should start for each sample.
ADDITIONAL INFORMATION
id7453, 2767, 2767SFC, FLYNXV41, UNIFSW17