ABPR Health Test failed on ACQUITY UPC2 - WKB72439
Article number: 72439
SYMPTOMS
- ABPR Health Test failed on ACQUITY UPC2.
- Minimum Pressure test fail with 98.2 bar pressure.
ENVIRONMENT
- ACQUITY UPC2
- Convergence Manager
CAUSE
Static back pressure regulator blocked.
FIX or WORKAROUND
Replace static back pressure regulator (700009459)
ADDITIONAL INFORMATION
id72439, UPC2BSM, UPC2CMA, UPC2CNVMGR, UPC2SMFL