SFC Prep 100 system overpressure - WKB64436

Article number: 64436

SYMPTOMS

  • SFC Prep 100 system gives overpressure errors and does not stabilize

ENVIRONMENT

  • SFC Prep 100
  • P-200 Pump
  • 2535/2545/2555 Quaternary Gradient Module

CAUSE

Blocked in-line filter

FIX or WORKAROUND

  1. Identify the filter that causes elevated pressure.
  2. Replace the in-line filter (700008142 - Filter, 2 Micron Iso-Prep Replacement).

ADDITIONAL INFORMATION