P-50A co-solvent pump does not deliver solvent at the set flow rate - WKB5884

Article number: 5884

SYMPTOMS

  • Waters P-50A co-solvent pump does not deliver solvent at the set flow rate
  • Set flow rate = 10 mL/min; measured flow rate = 2 mL/min
  • No flow from P50X
  • Solvent and tube are normal
  • RPM of the pump is in range

ENVIRONMENT

  • Waters P-50A co-solvent pump
  • Thar P-50A co-solvent pump
  • SFE Bio-Botanical Extraction System
  • SFC 80

CAUSE

The check valves are worn or have failed and need replacement.

FIX or WORKAROUND

  1. Inspect to see if the check valves are blocked; try to ultrasonic them.
  2. If step 1 has no effect, order four new check valves (part 700006600 for 1).
  3. Replace the check valves. 
  4. Verify that the issue is resolved.