How to get a 1525 pump to start in Empower - WKB57296

Article number: 57296

OBJECTIVE or GOAL

In Empower, start a 1525 pump, which is to be used only to flush the column.

ENVIRONMENT

  • 1525 Binary HPLC Pump
  • Empower 3

PROCEDURE

A chromatographic system that includes only the pump will not start because there is no "trigger" to start the run. To get the pump to work:

  1. Configure an optical detector.
  2. Create a new instrument method and chromatographic system. 

ADDITIONAL INFORMATION

With a pump-only configuration, within Run Sample Set, the sample set cannot be run, and the alert “system idle, sample set run finish” then appears.

id57296, 1525, 1525U, 1525UIVD