Low pulsing pressure on 2545 Binary Gradient Module - WKB56087
Article number: 56087
SYMPTOMS
- Pulsing pressure lower than usual
- High delta psi
ENVIRONMENT
- MassLynx
- Autopurification
- 2525/2545 Binary Gradient Module
CAUSE
Failed seals on a pump
FIX or WORKAROUND
- Run the leak test to identify which pump is the problem.
- Rebuild the pump with replacement seals.
ADDITIONAL INFORMATION
Associated part numbers:
Part Number 700001494 - Kit, Plunger Seal Kit, PKG 4, BGM which includes:
Part Number 425000204 - Seal, Wash Housing, Floating Flanged
Part Number 425000248 - Flanged Seal, .280 ID, High Pressure
Part Number 425000270 - O-Ring, Teflon, 1.130 OD X .103 Wide, 2-1