The view cell on the SFE Bio-Botanical Extraction System is leaking - WKB53501
Article number: 53501
SYMPTOMS
- CO2 is coming out of the exhaust port of the flow cell
ENVIRONMENT
- SFE Bio-Botanical Extraction System
CAUSE
Warped or damaged O-ring
FIX or WORKAROUND
Replace the O-rings (700006830) on the inside of the view cell.
ADDITIONAL INFORMATION

There is an O-ring on each side. If you want to replace both O-rings, you must order two.