Baseline ripple in 2998 SFC PDA data - WKB51390

Article number: 51390

SYMPTOMS

  • Baseline ripple in the PDA-acquired chromatography 
  • Flow cell leaks

clipboard_e28406eb14c772ac83da376339813c862.pngclipboard_ef896e7a720742ca80eb77a11e0174e30.png

ENVIRONMENT

  • SFC 2998 PDA

CAUSE

An analytical flow cell rated 1000 psi (70 bar), instead of an SFC PDA flow cell rated 5800 psi (400 bar), was installed on an SFC 2998 PDA. This caused high pressure, which resulted in a leak in the detector flow cell.

FIX or WORKAROUND

Replace the cell assembly (700008232 - Cell Assy, 2998 Analyt 10 mm PL-400 Bar).

 

ADDITIONAL INFORMATION

 

id51390, 2998, 2998B, 2998SFC, A-10PDA, baseline noise