"Low Gas" error on 2424 - WKB50736

Article number: 50736

SYMPTOMS

  • Low Gas error

ENVIRONMENT

  • 2424
  • MassLynx

CAUSE

Nitrogen inlet pressure (set at 80 psi) is too low for the operation.

FIX or WORKAROUND

Set the nitrogen inlet pressure to 90-100 psi.

ADDITIONAL INFORMATION