Skip to main content
Waters

ACQUITY UPC2 Convergence Manager ABPR power test fails - WKB29945

Article number: 29945

SYMPTOMS

  • ABPR power test fails
  • ABPR minimum pressure test passes
  • No leaks found

ENVIRONMENT

  • ACQUITY UPC2 Convergence Manager

CAUSE

The ABPR has failed.

FIX or WORKAROUND

  1. Replace the ABPR (700009051).

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.