Pressure drops with 2545 QGM - WKB290045

Article number: 290045

SYMPTOMS

  • Random pressure drops      

ENVIRONMENT

  • 2545 Quaternary Gradient Module

CAUSE

Back pressure is too low, sporadically causing check valves to not close properly. 

FIX or WORKAROUND

Install pressure-generating pre-column tubing to increase back pressure.

ADDITIONAL INFORMATION