High pressure on SFC Prep 150 system - WKB272578
Article number: 272578
SYMPTOMS
- High pressure on SFC Prep 150 system
ENVIRONMENT
- SFC Prep 150
CAUSE
The outlet tubing from the column oven valve to the splitter is blocked, as is the splitter inline filter.
FIX or WORKAROUND
Replace the stainless tubing (part number WAT026805) and filter (part number 700003269) as appropriate.