2707 Sample Manager leaks from needle during aspiration - WKB267545

Article number: 267545

SYMPTOMS

  • 2707 Sample Manager leaks from needle during aspiration

ENVIRONMENT

  • 2707

CAUSE

The rotor seal is damaged.

FIX or WORKAROUND

Replace the rotor seal (for analytical system, part 700003851; for prep system, part 700005097).

ADDITIONAL INFORMATION