Autopurification system hangs waiting on inject start - WKB257936
Article number: 257936
SYMPTOMS
- Autopurification system hangs waiting on inject start
- 2767 injects sample, but acquisition fails to start
ENVIRONMENT
- 2767
- Autopurification
- QDa
- ELSD
- PDA
- 2998
- 2424
- MassLynx
CAUSE
Inject start cables incorrectly connected on the back of detector or 2767
FIX or WORKAROUND
Correctly fit the inject start cables to the inject start positions for the specific detector or injector.
ADDITIONAL INFORMATION